The ACE1 Electrical Impedance Tomography System for Thoracic Imaging
- Michelle M. Mellenthin,
- Jennifer L. Mueller(corresponding author),
- Erick Dario León Bueno de Camargo,
- Fernando Silva de Moura,
- Talles Batista Rattis Santos,
- Raul Gonzalez Lima
- Colorado State University,
- University of Colorado,
- Federal University of ABC,
- University of São Paulo,
- Marquette University,
- Rowan University
Abstract
The design and performance of the active complex electrode (ACE1) electrical impedance tomography system for single-ended phasic voltage measurements are presented. The design of the hardware and calibration procedures allows for reconstruction of conductivity and permittivity images. Phase measurement is achieved with the ACE1 active electrode circuit which measures the amplitude and phase of the voltage and the applied current at the location at which current is injected into the body. An evaluation of the system performance under typical operating conditions includes details of demodulation and calibration and an in-depth look at insightful metrics, such as signal-to-noise ratio variations during a single current pattern. Static and dynamic images of conductivity and permittivity are presented from ACE1 data collected on tank phantoms and human subjects to illustrate the system’s utility.
Bibliographic Information
Output type
Original language
EnglishPages from-to (Number of pages)
Pages 3137-3150 (14 pages)Journal (Volume, Issue Number)
IEEE Transactions on Instrumentation and Measurement (Volume 68, Issue 9)Publication milestones
- Published - 01/09/2019
Publication status
ISSN
0018-9456Publication IDs
- Scopus: 85047369775